Associated content: 

In Situ Measurements of Aerosol Microphysical Properties

Two instruments,a nucleation-mode aerosol size spectrometers (NMASS; Williamson et al., 2018), and an ultra-high sensitivity aerosol spectrometers (UHSAS; Kupc et al., 2018) comprise the AMP package for ACCLIP. The AMP package provides particle size distributions with up to one-second time resolution for dry aerosol particles between 0.003 and 1.5 µm in diameter. Details of methods, uncertainties, and data products from the AMP package are in Brock et al. (2019).

During ATom, the instruments were used to investigate how particles in the remote atmosphere influence climate by examining the origin of small particles in the remote atmosphere and their growth to sizes where they can affect clouds and the sources, characteristics, and distribution of soil dust and sea-spray particles, and 3) the importance long-range transport from human and natural sources on background aerosol properties.

Instrument Type: 
Aircraft: 
DC-8 - AFRC, Gulfstream G-5
Point(s) of Contact: 

Programmable Flask Package Whole Air Sampler

The PFP whole air sampler provides a means of automated or manual filling of glass flasks, twelve per PFP. The sampler is designed to remove excess water vapor from the sampled air and compress it without contamination into ~1-liter volumes. These flasks are analyzed at the NOAA’s Global Monitoring Division laboratory for trace gasses and at  the INSTAR’s Staple Isotope Lab laboratory for isotopes of methane. More than 60 trace gases found in the global atmosphere can be measured at mole fractions that range from parts-per-million (10-6), e.g., carbon dioxide, down to parts-per-quadrillion (10-15), e.g., HFC-365mfc.  The chemical species monitored include N2O, SF6, H2, CS2, OCS, CO2, CH4, CO, CFCs, HCFCs, HFCs, Solvents, Methyl Halides, Hydrocarbons and Perfluorocarbons.

Instrument Type: 
Aircraft: 
Point(s) of Contact: 

In Situ Airborne Formaldehyde

The NASA GSFC In Situ Airborne Formaldehyde (ISAF) instrument measures formaldehyde (CH2O) on both pressurized and unpressurized (high-altitude) aircraft. Using laser induced fluorescence (LIF), ISAF possesses the high sensitivity, fast time response, and dynamic range needed to observe CH2O throughout the troposphere and lower stratosphere, where concentrations can range from 10 pptv to hundreds of ppbv.

Formaldehyde is produced via the oxidation of hydrocarbons, notably methane (a ubiquitous greenhouse gas) and isoprene (the primary hydrocarbon emitted by vegetation). Observations of CH2O can thus provide information on many atmospheric processes, including:
 - Convective transport of air from the surface to the upper troposphere
 - Emissions of reactive hydrocarbons from cities, forests, and fires
 - Atmospheric oxidizing capacity, which relates to formation of ozone and destruction of methane
In situ observations of CH2O are also crucial for validating retrievals from satellite instruments, such as OMI, TROPOMI, and TEMPO.

Instrument Type: 
Measurements: 
Aircraft: 
Point(s) of Contact: 

In-Situ Measurements of Aerosol Optical Properties

Three instruments, a cavity ringdown (CRD) aerosol extinction spectrometer, a photoacoustic absorption spectrometer (PAS), and an ultra-high sensitivity aerosol size spectrometer (UHSAS) comprise the AOP package. The AOP package provides multi‐wavelength, multi-RH aerosol extinction and absorption measurements with fast response and excellent accuracy and stability on aircraft platforms. The instruments will also characterize the optics of black carbon (BC) mixing state, brown carbon, and water uptake of aerosol. Aerosol asymmetry parameter, needed for radiative transfer modeling, will be calculated from dry and humidified particle size distributions.

Instrument Type: 
Point(s) of Contact: 

NOAA Nitrogen Oxides and Ozone

The NOAA NOyO3 4-channel chemiluminescence (CL) instrument will provide in-situ measurements of nitric oxide (NO), nitrogen dioxide (NO2), total reactive nitrogen oxides (NOy), and ozone (O3) on the NASA DC-8 during the FIREX-AQ project. Different versions of this instrument have flown on the NASA DC-8 and NOAA WP-3D research aircraft on field projects since 1995. It provides fast-response, specific, high precision, and calibrated measurements of nitrogen oxides and ozone at a spatial resolution of better than 100m at typical DC-8 research flight speeds. Detection is based on the gas-phase CL reaction of NO with O3 at low pressure, resulting in photoemission from electronically excited NO2. Photons are detected and quantified using pulse counting techniques, providing ~5 to 10 part-per-trillion by volume (pptv) precision at 1 Hz data rates. One detector of the integrated 4-channel instrument is used to measure ambient NO directly, a second detector is equipped with a UV-LED converter to photodissociate ambient NO2 to NO, and a third detector is equipped with a heated gold catalyst to reduce ambient NOy species to NO. Reagent ozone is added to these sample streams to drive the CL reactions with NO. Ambient O3 is detected in the fourth channel by adding reagent NO.

Instrument Type: 
Measurements: 
Aircraft: 
Point(s) of Contact: 

CU Aircraft High-Resolution Time-of-Flight Aerosol Mass Spectrometer

Principle: The CU aircraft version of the Aerodyne High-Resolution Time-of-Flight Aerosol Mass Spectrometer (HR-ToF-AMS) detects non-refractory submicron aerosol composition by impaction on a vaporizer at 600°C, followed by electron ionization and time-of-flight mass spectral analysis. Size-resolved composition can be quantified by measuring the arrival times of the aerosol at the vaporizer.

Aircraft Operation: (1 min cycles, can be adjusted to meet mission goals):
46 s total concentration measurements (1 s resolution, can be increased to up to 10 Hz upon request)
5 s speciated size distribution measurements (with improved S/N detection due to ePToF acquisition)
9 s Background + Overhead
Higher accuracy due to flight day calibrations using built-in system
Custom pressure controlled inlet with confirmed performance up to 45 kft

Real Time Data Products: 
PM1 Aerosol Mass Concentrations:
Organic aerosol (OA) , SO4, NO3, NH4, Chloride 
OA Chemical Markers: f44 (Secondary OA), f57 (hydrocarbon-like OA), f60 (biomass burning OA), f82 (isoprene epoxide-SOA), other fx upon request

More Advanced Products:
- PM1 Seasalt, ClO4, total I, total Br, MSA concentrations
- O/C, H/C, OA/OC, OSc
- Particle organic nitrates (pRONO2)
- Ammonium Balance, estimated pH
- OA components by positive matrix factorization (PMF)
- Particle eddy covariance fluxes of all species
- Speciated Aerosol size distributions

Detection Limits (1s, ng sm-3), (1 min, ng sm-3) from start of the flight (due to custom cryopump):
Sulfate: 40, 15
Nitrate: 15, 6
Ammonium: 3, 1
Chloride: 30, 12
OA: 200, 80
For detailed OA analysis, longer averaging (3-30 s, depending on OA concentration) is needed. A 1 min product is hence provided as well.

 

Instrument Type: 
Aircraft: 
Point(s) of Contact: 

TSI 3563 Integrating Nephelometer

TSI Integrating Nephelometers are designed specifically for studies of direct radiative forcing of the Earth’s climate by aerosol particles, or studies of ground-based or airborne atmospheric visual air quality in clean areas. They may also be used as an analytical detector for aerosol particles whenever the parameter of interest is the light-scattering coefficient of the particles after a pretreatment step, such as heating, humidification, or segregation by size. The light-scattering coefficient is a highly variable aerosol property. Integrating Nephelometers measure the angular integral of light scattering that yields the quantity called the aerosol scattering coefficient, which is used in the Beer-Lambert Law to calculate total light extinction.

Measurements: 

Particle Into Liquid Sampler

The Particle Into Liquid Sampler (PILS) was developed for rapid automated on-line and continuous measurement of ambient aerosol bulk composition. The general approach is based on earlier devices in which ambient particles are mixed with saturated water vapor to produce droplets easily collected by inertial techniques. The resulting liquid stream is analyzed with an ion chromatograph to quantitatively measure the bulk aerosol ionic components. In this instrument, a modified version of a particle size magnifier is employed to activate and grow particles comprising the fine aerosol mass. A single jet inertial impactor is used to collect the droplets onto a vertical glass plate that is continually washed with a constant water diluent flow of nominally 0.10 ml min-1. The flow is divided and then analyzed by a dual channel ion chromatograph. In its current form, 4.3 min integrated samples were measured every 7 min. The instrument provides bulk composition measurements with a detection limit of approximately 0.1 µg m-3 for chloride, nitrate, sulfate, sodium, ammonium, calcium, and potassium.

Instrument Type: 
Measurements: 
Na, NH4, K, Mg, Ca+2, Cl, NO2, NO3, SO4, PO4, Br-, WSOC
Point(s) of Contact: 

Single Particle Soot Photometer (NOAA)

The SP2 is a laser-induced incandescence instrument primarily used for measuring the refractory BC  (rBC) mass content of individual accumulation-mode aerosol particles. It is able to provide this data product independently of the total particle morphology and mixing state, and thus delivers detailed information not only about BC loadings, but also size distributions, even in exceptionally clean air. The instrument can also provide the optical size of individual particles containing rBC, and identify the presence of materials associated with the BC fraction (i.e. identify the rBC’s mixing state). Since its introduction in 2003, the SP2 has been substantially improved, and now can be considered a highly competent instrument for assessing BC loadings and mixing state in situ.  NOAA deploys multiple SP2s with different designs: the first was built for the WB-57F research aircraft. Two others are rack-mounted units customized at NOAA; one of the rack mounted units can be humidified, and has been deployed with a paired dry rack-mounted SP2 as the "Humidified-Dual SP2" (HD-SP2). The rack mounted units are suitable for in-cabin operations.

Instrument Type: 
Measurements: 
Point(s) of Contact: 

Scanning High-Resolution Interferometer Sounder

The Scanning High-resolution Interferometer Sounder (S-HIS) is a scanning interferometer which measures emitted thermal radiation at high spectral resolution between 3.3 and 18 microns The measured emitted radiance is used to obtain temperature and water vapor profiles of the Earth's atmosphere in clear-sky conditions. S-HIS produces sounding data with 2 kilometer resolution (at nadir) across a 40 kilometer ground swath from a nominal altitude of 20 kilometers onboard a NASA ER-2 or Global Hawk.

Instrument Type: 
Measurements: 
Point(s) of Contact: 

Pages

Subscribe to RSS - AEROMMA