Differential Aerosol Sizing and Hygroscopicity Spectrometer Probe

The DASH-SP providse rapid measurements of size-resolved aerosol sub-saturated hygroscopic growth factors and the real part of aerosol refractive index. It has been deployed aboard the NASA DC-8 during the DC3 and SEAC4RS field campaigns.

Measurements: 
Point(s) of Contact: 

G-V Scanning Mobility Particle Sizer

The GV Scanning Mobility Particle Sizer (SMPS) measures the particle size distribution over the mobility diameter range of 3 to 500 nm (pressure-dependent). It consists of two components: an electrostatic classifier (EC) and a condensation particle counter (CPC). The EC samples aerosol-laden ambient air, places a well-defined charge distribution on the particles, and then selects a narrow range of particle “mobility diameter” (approx. equal to cross-sectional area-to-charge ratio) using a differential mobility analyzer (DMA). The selected diameter can be scanned by a time-varying high voltage applied to the DMA; following this particles are counted by the CPC. The total scan time and the number of counting intervals, the latter of which determines the number of diameter bins in the size distribution, are selected based on ambient particle concentrations and altitude. The raw data (particle counts over each counting interval as a function elapsed time during the linear diameter scan) is mathematically inverted during post-processing to obtain the particle size distribution.

Instrument Type: 
Aircraft: 
Gulfstream V - NSF
Point(s) of Contact: 

Hawaii Group for Environmental Aerosol Research

1) Time of Flight Aerosol Mass Spectrometer (ToF-AMS)

Total and single particle characterization of volatile aerosol ionic and organic components (50-700nm). Uncertainty depends on species and concentration.

2) Single Particle Soot Photometer (SP2)

Single particle measure of BC (soot) mass in particles and determination of mixed particle size and non-BC coating using laser scattering and incandescence. 70-700nm. Single particle counting up to 10,000 per sec.

3) A size-resolved thermo-optic aerosol discriminator (30 s avg.):

Aerosol size distribution from 0.12 up to 7.0 μm, often where most aerosol mass, surface area and optical effects are dominant. Uses a modified Laser Optical Particle Counter (OPC) and computer controlled thermal conditioning system is used upstream (airstream dilution dried). Characterizes aerosol components volatile at 150, 300 and 400C and refractory aerosol at 400C (sea salt, dust and soot/flyash). (Clarke, 1991, Clarke et al., 2004). Uncertianty about 15%

4) Condensation Nuclei - heated and unheated (available at 1Hz)

Two butanol based condensation nuclei (CN) counter (TSI 3010) count all particles between 0.01-3.0 um. Total CN, refractory CN (those remaining at 300C after sulfate is removed) and volatile CN (by difference) are obtained as a continuous readout as a fundamental air mass indicator (Clarke et al. 1996). Uncertainty ~ 5%.

5) Aerodynamic Particle Sizer – (APS-TSI3320) – (<5min/scan)

To further characterize larger “dry” particles, including dust, an APS is operated which sizes particles aerodynamically from 0.8 to 20 μm into 50 channels. Uncertainty~10%.

6) Differential Mobility Analyzer with thermal conditioning – (<3 min/scan)

Volatility tandem thermal differential mobility analyzer (VTTDMA) with thermal analysis that provides size information (mass, surface area, number distributions) and their state of mixing over the 0.01 to 0.3μm size range (Clarke et al., 1998, 2007) for sampling times of about 1-3 minutes. Uncertainty ~10%

7) Nephelometer (10-7 m-1 detection for 60s avg., recorded every 1 sec.)

A 3 wavelength nephelometer (450, 550, 700nm) is used for total scattering and submicrometer scattering values using a Radiance Research single wavelength nephelometer (and thereby coarse dust scattering by difference).

8) Two Particle Soot Absorption Photometers (PSAP-Radiance Research; detection <0.1μg m-3 for 5 min. avg. )

The PSAP is used to quantify the spectral light absorption coefficient of the total and submicron aerosol (eg. soot, BC) at three wavelengths (450, 550, 660nm).

9) Humidity Dependent Light-Scattering (10-6 m-1 detection for 60s avg.; recorded every 1 s)

Two additional Radiance Research single-wavelength nephelometers are operated at two humidities (high/low) to establish the humidity dependence of light scattering, f(RH).

Point(s) of Contact: 