HCFCs

Programmable Flask Package Whole Air Sampler

The PFP whole air sampler provides a means of automated or manual filling of glass flasks, twelve per PFP. The sampler is designed to remove excess water vapor from the sampled air and compress it without contamination into ~1-liter volumes. These flasks are analyzed at the NOAA’s Global Monitoring Division laboratory for trace gasses and at  the INSTAR’s Staple Isotope Lab laboratory for isotopes of methane. More than 60 trace gases found in the global atmosphere can be measured at mole fractions that range from parts-per-million (10-6), e.g., carbon dioxide, down to parts-per-quadrillion (10-15), e.g., HFC-365mfc.  The chemical species monitored include N2O, SF6, H2, CS2, OCS, CO2, CH4, CO, CFCs, HCFCs, HFCs, Solvents, Methyl Halides, Hydrocarbons and Perfluorocarbons.

Instrument Type
Aircraft
Point(s) of Contact
Advanced Whole Air Sampler

32 samples/flight (ER-2); 50 samples/flight (WB57); 90 samples/flight (Global Hawk)

Updated control system with remote control capability

Fill times
–14 km 30 – 40 sec
–16 km 40 – 50 sec
–18 km 50 – 60 sec
–20 km 100 – 120 sec (estimated)

Analysis in UM lab: GC/MS; GC/FID; GC/ECD

Instrument Type
Point(s) of Contact