Submillimeter Limb Sounder

The Submillimeterwave Limb Sounder (SLS) is a heterodyne radiometer measuring thermal emission spectra near 640 GHz (for detection of ClO, HCl, and O3) and 604 GHz. (for detection of HNO3 and N2O) designed for use on high altitude balloons and aircraft. The instrument consists of five subsystems:

-optics which define the instrument field of view (FOV)
-radiometer front-ends which down converts incoming radiance signals
-intermediate frequency (IF) stage which selects and frequency shifts signal bands
-spectrometers which frequency resolve and detect the incoming power spectrum
-command and data handling which controls the instrument and transmits data to the ground

Limb scanning is accomplished by a flat mirror (~20 cm diameter) connected to a stepper motor (0.2 steps) and 14 bit position encoder. This mirror is also used for gain and zero calibration by viewing an absorber target located below the mirror and upward at 47° elevation angle to view the cold sky. A set of three off-axis parabolic reflectors form the instrument field of view (0.35 full width at half maximum) and couple limb radiance to the mixer input waveguide. These reflectors are oversized (~30 dB edge taper) to minimize side lobes in the FOV. Pointing and beam shape were verified by scanning the instrument FOV across the emission from a 600 GHz transmitter (multiplied output of a Gunn oscillator) located in the receiver optical far-field.

The radiometer front-end is an uncooled second harmonic mixer using a waveguide mounted Schottky diode. The radiometer is operated double side band (DSB), i.e., spectral features occurring symmetrically above and below the effective local oscillator frequency (637.050 GHz) appear together in the IF output spectrum. The diode is pumped at a 318.525 GHz. This source is generated by a tripled 106.175 GHz phase-locked InP Gunn oscillator and wave guide coupled to the mixer block. The mixer produces an IF output spectrum of 10.5 to 13 GHz, which corresponds to signals at the mixer input at 647.5 GHz to 650.0 GHz (in the radiometer upper side band) and 626.5 GHz to 624.1 GHz ( in the lower side band). The design of the 604 GHz radiometer system is similar to 637 GHz system but operates at a lower IF frequency of 2 to 3 GHz.

Diagram of the SLS frequency down-conversion scheme. RF signals enter the signal flow path through mixer feeds at the left of the diagram. At the right side, the signal flow enters a set of UARS MLS-type filterbank spectrometers where bands are further spectrally resolved, power detected, and digitized.

Instrument Type: 
Measurements: 
Aircraft: 
Balloon, ER-2 - AFRC
Point(s) of Contact: 

Dual-Beam UV-Absorption Ozone Photometer

The NOAA-O3 instrument consists of a mercury lamp, two sample chambers that can be periodically scrubbed of ozone, and two detectors that measure the 254-nm radiation transmitted through the chamber. The ozone absorption cross-section at this wavelength is accurately known; hence, the ozone number density can be easily calculated. Since the two absorption chambers are identical, virtually continuous measurements of ozone are made by alternating the ambient air sample and ozone scrubbed sample between the two chambers. At a one-second data collection rate, the minimum detectable concentration of ozone (one standard deviation) is 1.5 x 10 10 molecules/cm 3 (0.6ppbv at STP).

Instrument Type: 
Measurements: 
Aircraft: 
Point(s) of Contact: 

JPL Mark IV Balloon Interferometer

The MkIV interferometer operates in solar absorption mode, meaning that direct sunlight is spectrally analyzed and the amount of various gases at different heights in the Earth's atmosphere is derived from the shapes and depths of their absorption lines. The optical design of the MkIV interferometer is based largely on that of the ATMOS instrument, which has flown four times on the Space Shuttle. The first three mirrors in the optical path comprise the suntracker. Two of these mirrors are servo-controlled in order to compensate for any angular motion of the observation platform. The subsequent wedged KBr plates, flats, and cube-corner retro-reflectors comprise a double-passed Michelson interferometer, whose function is to impart a wavelength-dependent modulation to the solar beam. This is achieved by sliding one of the retro-reflectors at a uniform velocity so that the recombining beams interfere with each other. A paraboloid then focusses the solar beam onto infrared detectors, which measure the interferometrically modulated solar signal. Finally, Fourier transformation of the recorded detector outputs yields the solar spectrum. An important advantage of the MkIV Interferometer is that by employing a dichroic to feed two detectors in parallel, a HgCdTe photoconductor for the low frequencies (650-1850 cm-1) and a InSb photodiode for the high frequencies (1850-5650 cm-1), the entire mid-infrared region can be observed simultaneously with good linearity and signal-to-noise ratio. In this region over 30 different gases have identifiable spectral signatures including H2O, O3, N2O, CO, CH4, NO, NO2, HNO3, HNO4, N2O5, H2O2, ClNO3, HOCl, HCl, HF, COF2, CF4, SF6, CF2ClCFCl2, CHF2Cl, CF2Cl2, CFCl3, CCl4, CH3Cl, C2H2, C2H6, OCS, HCN, N2, O2, CO2 and many isotopic variants. The last three named gases, having well known atmospheric abundances, are important in establishing the observation geometry of each spectrum, which otherwise can be a major source of uncertainty. Similarly, from analysis of T-sensitive CO2 lines, the temperature profile can be accurately determined. The simultaneity of the observations of all these gases greatly simplifies the interpretation of the results, which are used for testing computer models of atmospheric transport and chemistry, validation of satellite data, and trend determination.

Although the MkIV can measure gas column abundances at any time during the day, the highest sensitivity to atmospheric trace gases is obtained by observing sunrise or sunset from a balloon. The very long (~ 400 km) atmospheric paths traversed by incoming rays in this observation geometry also make this so-called solar occultation technique insensitive to local contamination.

Instrument Type: 
Aircraft: 
Balloon, DC-8 - AFRC
Point(s) of Contact: 

High-Sensitivity Fast-Response CO2 Analyzer

The high-sensitivity fast response CO2 instrument measures CO2 concentrations in situ using the light source, gas cells, and solid-state detector from a modified nondispersive infrared CO2 analyzer (Li-Cor, Inc., Lincoln, NE). These components are stabilized along the detection axis, vibrationally isolated, and housed in a temperature-controlled pressure vessel. Sample air enters a rear-facing inlet, is preconditioned using a Nafion drier (to remove water vapor), then is compressed by a Teflon diaphragm pump. A second water trap, using dry ice, reduces the sample air dewpoint to less than 70C prior to detection. The CO2 mixing ratio of air flowing through the sample gas cell is determined by measuring absorption at 4.26 microns relative to a reference gas of known concentration. In-flight calibrations are performed by replacing the air sample with reference gas every 10 minutes, with a low-span and a high-span gas every 20 minutes, and with a long-term primary standard every 2 hours. The long-term standard is used sparingly and serves as a check of the flight-to-flight accuracy and precision of the measurements, augmented by ground-based calibrations before and after flights.

Instrument Type: 
Measurements: 
Aircraft: 
Point(s) of Contact: 

Lightweight Balloon Interferometer

The LWBI includes an optical head, containing an interferometer, detectors, translation mechanism, and metrology laser, using light weight technologies and a more mass-conscious approach. The lightweight balloon interferometer (LWBI) will have a performance similar to that of the JPL MkIV, yet will weigh just one fifth of the mass. The reduced payload size and mass will make the LWBI much easier to launch, facilitating close co-location with observations from space-borne sensors such as TES and HIRDLS on board AURA.

Instrument Type: 
Aircraft: 
Balloon
Point(s) of Contact: 

CO2 Weather Balloon Spectrometer

Instrument Type: 
Measurements: 
Aircraft: 
Balloon
Point(s) of Contact: 

Continuous Flow Streamwise Thermal Gradient CCN Counter

Developed by Droplet Measurement Technologies, the CFSTGC is based on a concept by Roberts and Nenes [2005]. The instrument counts the fraction of aerosol particles that become droplets when exposed to a given water vapor supersaturation (RH > 100%).

As with all CCN counters, a temperature gradient is applied to produce a supersaturation of water vapor. However, the mechanism for generating supersaturation is not the same for all CCN counters. For example, for continuous flow parallel plate diffusion chambers, the temperature gradient is perpendicular to the flow, and supersaturation is a result of the nonlinear dependence of vapor pressure upon temperature. The same mechanism applies for static diffusion cloud chambers, where there is no flow at all.

However, as the name implies, for the Continuous Flow Streamwise Thermal Gradient CCN Counter, the temperature gradient is in the streamwise direction (maintained by thermoelectric coolers). In this case, supersaturation results as a consequence of the greater rate of mass transfer over heat transfer.

With laminar flow, heat and water vapor are transferred to the centerline of the column from the walls only by diffusion.

Since molecular diffusivity is greater than thermal diffusivity, the distance downstream that a water molecule travels before reaching the centerline is less than the distance the heat travels downstream before reaching the centerline. If you pick a point at the centerline, the heat originated from a greater distance upstream than the water vapor.

There are four facts that are necessary to explain how supersaturation is generated within the CFSTGC:

1) Assuming that the inner surface of the column is saturated with water vapor at all points, since the temperature is greater at point B than at point A, the water vapor partial pressure is also greater at point B than at point A.

2) The actual partial pressure of water vapor at point C is equal to the partial pressure of water vapor at point B.

3) However, since the temperature at point C is the same as at point A, the equilibrium water vapor pressure at point C is equal to the water vapor partial pressure at point A.

4) The saturation ratio is the ratio between the actual partial pressure of water vapor and the equilibrium vapor pressure. This is equivalent to the partial pressure at point B divided by the partial pressure at point A, which is always greater than one. Thus supersaturation is generated through a dynamic equilibrium.

Instrument Type: 
Measurements: 
Aircraft: 
Balloon, P-3 Orion - WFF, C-130H - WFF, DC-8 - AFRC, HU-25 Falcon - LaRC
Point(s) of Contact: 

Argus Tunable Diode Laser Instrument

Argus is a two channel, tunable diode laser instrument set up for the simultaneous, in situ measurement of CO (carbon monoxide), N2O (nitrous oxide) and CH4 (methane) in the troposphere and lower stratosphere. The instrument measures 40 x 30 x 30 cm and weighs 21 kg. An auxiliary, in-flight calibration system has dimensions 42 x 26 x 34 cm and weighs 17 kg.

The instrument is an absorption spectrometer operating in rapid scan, secondharmonic mode using frequency-modulated tunable lead-salt diode lasers emitting in the mid-infrared. Spectra are co-added for two seconds and are stored on a solid state disk for later analysis. The diode laser infrared beam is shaped by two anti-refection coated lenses into an f/40 beam focused at the entrance aperture of a multi-pass Herriott cell. The Herriott cell is common to both optical channels and is a modified astigmatic cell (New Focus Inc., Santa Clara, California).

The aspherical mirrors are coated with protected silver for optimal infrared reflectivity. The cell is set up for a 182-pass state for a total path of 36m. The pass number can be confirmed by visual spot pattern verification on the mirrors observed through the glass cell body when the cell is illuminated with a visible laser beam. However, instrument calibration is always carried out using calibrated gas standards with the Argus instrument operating at its infrared design wavelengths, 3.3 and 4.7 micrometers respectively for CH4 and CO detection. The electronic processing of the second harmonic spectra is done by standard phase sensitive amplifier techniques with demodulation occurring at twice the laser modulation frequency of 40 kHz. To optimize the secondharmonic signal amplitude in a changing ambient pressure environment the laser modulation amplitude is updated every 2 seconds to its optimal theoretical value based upon the measured pressure in the Herriott cell.

Measurements: 
Aircraft: 
Point(s) of Contact: 

Aircraft Laser Infrared Absorption Spectrometer

ALIAS (Aircraft Laser Infrared Absorption Spectrometer) measures total water, total water isotopes, carbon monoxide, and carbon dioxide isotope ratios. No other instrument provides real-time measurements of carbon dioxide isotope ratios which are clear identifiers of atmospheric transport (18O/17O/16O for stratospheric intrusion, 13C/12C for anthropogenic signals). ALIAS easily adapts to changing mission priorities and can be configured to measure HCl, CH4, SO2, and N2O by simply replacing a semiconductor laser. These measurements contribute to Atmospheric Composition Focus Area research by providing key data on how convective processes affect stratospheric composition, the development of cirrus particles and their affect on Earth's radiative balance, and health of the ozone layer through measurement of chlorine partitioning.

Measurements: 
Aircraft: 
Point(s) of Contact: 

Airborne 2-Channel Laser Infrared Absorption Spectrometer

The Airborne Laser Infrared Absorption Spectrometer (ALIAS-II) is a very high resolution scanning tunable diode laser spectrometer which makes direct, simultaneous measurements of selectable combinations of HCl, NO2, CO, CO2, CH4, and N2O at sub-part-per-billion levels over a 3-30 second integration time. The measurement technique is based upon using tunable lead-salt and/or quantum cascade lasers operating from 3.4 to 8 microns wavelength scanning over absorption lines at 10 Hz recording second harmonic spectra. The instrument features an open-cradle multipass Herriott absorption cell with 15.24-cm diameter spherical zerodur mirrors coated with gold on chrome. The separation between the mirrors is adjustable allowing for a relatively small cell (0.75-m to 1.5-m) to contain an optical path length up to 120-m, depending on the spacing of the mirrors. Lasers and detectors are contained in a lightweight aluminum liquid nitrogen Dewar which can achieve a 28-hour hold time with only a 2 liter charge of liquid nitrogen. The instrument features custom laser current drives, signal chains, InSb detectors and preamps, 16-bit signal averager, analog signal conditioner, and digital I/O which are controlled by an onboard Pentium processor. Data is written to a ruggedized 2-Gb hard disk every 30 seconds and simultaneously transmitted via telemetry to ground station computers which provide backup storage of the data. The instrument weighs 36 kg and requires <56 watts for operation. Additional power up to 250 watts is available for structural heaters and current draw varies with atmospheric conditions.

Instrument Type: 
Measurements: 
Aircraft: 
Balloon
Point(s) of Contact: 

Pages

Subscribe to RSS - Balloon