The website is undergoing a major upgrade that began Friday, October 11th at 5:00 PM PDT. The new upgraded site will be available no later than Monday, October 21st. Until that time, the current site will be visible but logins are disabled.

PeroxyAcetylNitrate, Aldehydes and Ketones (PANAK)

Status

Operated By: 
PI

Don Hearlth, Rick Kolyer

The Ames PANAK instrument is a computerized 3- channel Capillary Gas Chromatographic system designed for the collection and analysis of low ppt (10-12 v/v) levels of peroxyacyl nitrates (PANs), alkyl nitrates, and tertrachloroethene in Channels 1 and 2; and C2-C3 aldehydes, C1-C2 alcohols, C3-C4 ketones, and C1-C2 nitriles in channel 3. Channels 1 and 2 use ECD detectors and have a sampling frequency of 2.5 minutes. Channel 3 uses a Photo Ionization detector placed in series with a Reduction Gas detector and has a sampling frequency of 5 minutes. The main manifold draws 5 SL/min of ambient air through a heated Teflon lined probe from which each of the three instrument channels draws a 200 ml aliquot of sample air. This aliquot is dried by passing it through a –35 °C cold trap, cooled to -140 °C for constituent pre concentration, and then heat desorbed into the gas chromatographic columns. All calibrations are performed in-flight by using an installed dilution system and in a manner that mimics ambient air sampling. Primary standards are generally referred to a series of permeation tubes. In addition high concentration standards are also carried on board. Sensitivities under typical conditions are: 1-3 ppt PANs, 1-5 ppt alkyl nitrates, 5-20 ppt OVOC, and 20-30 ppt nitriles.

Instrument Type: 
Aircraft: 
Point(s) of Contact: 
Hanwant Singh (POC; PI)